Karl Fischer Moisture Meter for Lithium Battery-AKF-CH7

Karl Fischer Moisture Meter for Lithium Battery-AKF-CH7

AKF – CH7  Karl Fischer moisture meter for is an integrated device combining a water – content – measurement module and a heating – sample – injection module. It has a novel appearance design and is convenient for use and maintenance. It can directly add liquid samples into the titration cell for quick moisture determination. It can also place insoluble samples in a sealed top – space sample bottle, heat the samples using the card – type heater equipped with the instrument, and then use an external dry carrier gas (nitrogen or air) to bring the water vapor evaporated from the sample bottle to the titration cell of the Karl Fischer moisture analyzer for determination.

Category:
Share your love
Share your love

Functional Features

Supports comprehensive detection of solid, liquid, and gas samples with accurate and reliable results.

Color touch – screen, full – digital keyboard, one – key start – up, simple operation.

Intelligent detection function: automatic balance, automatic detection, and automatic adjustment of detection precision in response to environmental changes.

Electrolysis control function: adapts to various electrolytes, with a high detection limit and a moisture – determination precision of up to 0.01ug.

Sample – bottle automatic – puncture lifting function: simple operation, effectively improving work efficiency.

Sample evaporator has multiple high – precision temperature – control modes such as constant temperature and constant rise, and is equipped with intelligent over – temperature protection.

Can edit, store, and call various analysis – method parameters to facilitate adaptation to different sample tests.

Built – in multiple measurement methods, simultaneously adapting to various detection requirements for gases, liquids, fixed samples, and low – water – content samples.

Intelligent status monitoring, hardware intelligent protection, abnormal termination, timing stop, etc., effectively extending product service life.

Technical Parameters

Power Supply Single – phase AC220V, Frequency 50Hz, Power approximately 300W
Gas Source Input pressure ≤ 0.6MPa, dry compressed air or dry nitrogen (99.99% purity)
Measurement Range 3ug – 199mg (H₂O mass)
Measurement Resolution 0.01ug (H₂O mass)
Measurement Repeatability > 99.7% (500ug pure water)
Sample Measurement Time Several tens of seconds to several tens of minutes (depending on the sample method)
Applicable Sample Types Liquids, solids, and others
Sample Injection Method Puncture injection, manual eight – position
Temperature Control Range Recommended 40 – 280°C / 0.1°C programmable
Heating Rate 15°C/min (below 200°C)
Flow Rate Range 0 – 100ml/min
Sample Bottle Specifications 10ml and 20ml
Data Output RS232/USB pin – type micro – printer
Equipment Size Approximately L460mm × W240mm × H360mm (subject to the actual product)