1100C 16-Channel Tube Furnace – GSL-1100X-MGI-16

1100C 16-Channel Tube Furnace – GSL-1100X-MGI-16

Application range:

GSL-1100X-MGI-16 is a 16-channel tube-type furnace, aiming at high-throughput sample sintering, annealing, and quenching operations, with a maximum temperature of 1100℃. Each heating module is controlled by an independent temperature-control system, and each furnace tube is sealed with vacuum flanges to heat-treat samples in vacuum or atmosphere-protected environments.

Model:

GSL-1100X-MGI-16

Dimension:

2350mm x 690mm x 1860mm (L x W x H); 900mm x 660mm x 1340mm (L x W x H)

Share your love

Features

  • The tube-type furnace has a total of 16 channels, and the size of each furnace tube is 25 OD × 20 ID × 200 L mm.
  • Each heating module has an independent temperature-control system, and the maximum temperature can reach 1100℃.
  • The quick-connect flange of KF25 is installed on each quartz tube, allowing samples to be heat-treated in vacuum or atmosphere-protected environments.
  • The heating module has two working positions. The horizontal position is used for sample sintering, and the vertical position is used for sample quenching. The furnace body can be rotated from the horizontal position to the vertical position, and can be used for metal sample pouring experiments.

Specifications

Working Temperature Continuous working temperature: 1000℃; maximum working temperature: 1100℃ (<1 hour)
Constant-Temperature Zone Length 30 mm (temperature uniformity: ±1℃); 60 mm (temperature uniformity: ±5℃)
Temperature-Control System The 16-channel has an independent temperature-control system, and all temperature settings are concentrated on a 6-inch touch screen for operation; the temperature is adjusted using the PID method, and 30-stage heating and cooling procedures can be set; it has over-temperature and disconnection protection programs
Maximum Heating Rate 10℃/min
Temperature Control Precision +/- 1℃; K-type thermocouples (16 in total) are inserted into each furnace tube for accurate temperature measurement and control
Furnace Tubes Quartz tubes (16 in total); dimensions: 25 OD × 20 ID × 200 L mm
Vacuum Flanges KF25 quick-connect flanges with mechanical pressure gauges installed on top
Maximum Sample Capacity The maximum sample capacity for each channel is 100g
Quenching Function The heating module has two working positions. The horizontal position is used for sample sintering, and the vertical position is used for sample quenching
Power & Operating Voltage Maximum power: 13KW; voltage: 208 – 240V single-phase, 50/60Hz
Vacuum Pumps (Optional) Various vacuum pumps are available for purchase from this company
Product Dimensions 2350mm L × 690mm W × 1860mm H; 900mm L × 660mm W × 1340mm H
Quality Certification All electrical components (>24V) have passed UL/MET/CSA/CE certification
Warranty One-year warranty and lifetime technical support (consumable parts such as heating elements, quartz tubes, and sample crucibles are not covered; damage caused by corrosive gases and acidic gases is not covered under the warranty)

Operational Notes

  • The air pressure inside the furnace tube should not exceed 0.02MPa.
  • Since the air pressure inside the gas cylinder is relatively high, a pressure-reducing valve must be installed on the gas cylinder when introducing gas into the furnace tube. It is recommended to purchase a pressure-reducing valve from our company. The pressure-reducing range of our company’s pressure-reducing valve is 0.01MPa – 0.1MPa, and the use will be more precise and safe.
  • When the furnace body temperature is higher than 1000℃, the furnace tube should not be in a vacuum state. The air pressure inside the furnace tube should be equivalent to the atmospheric pressure and maintained at a normal pressure state.
  • The gas flow rate into the furnace tube should be less than 200SCCM to avoid the impact of cold air flow on the heating quartz tube.
  • The long-term operating temperature of the quartz tube is <1100℃.
  • For sample heating experiments, it is not recommended to close the exhaust valve and inlet valve of the furnace tube flange. If it is necessary to close the valve for sample heating, pay close attention to the reading of the pressure gauge at all times. If the pressure reading is greater than 0.02MPa, the exhaust valve must be opened immediately to prevent accidents (such as furnace tube rupture, flange blow-out, etc.).