Functional Features
Cyclic voltammetry and linear sweep: Current averaging, controlled – current method, ohmic drop correction, corrosion rate analysis, and Tafel curve analysis.
Chronoamperometry, chronopotentiometry, and chronocoulometry: 1 to 255 different potentials/current values and ohmic drop compensation can be set, allowing for at least 65535 cycles of measurement.
Mixed mode: At least 255 segments can define applied signals, including potential steps, current steps, potential sweeps, current sweeps, open – circuit potential, controlled resistance and resistance sweeps; real – time impedance measurement from 10 Hz to 250 kHz; all steps and sweeps can monitor resistance and capacitance simultaneously during potential and current measurements; the entire process can be finite or continuously repeated.
Electrochemical noise: With electrochemical noise measurement and analysis functions.
AC impedance measurement and analysis functions: Can perform controlled – potential or controlled – current frequency sweeps; Mott – Schottky curves; electrochemical frequency modulation EFM; frequency range from 10 μHz to 250 kHz.
Photoelectrochemical functions: Through software’s I~V curve fitting function, short – circuit current (Isc), open – circuit voltage (Eoc), maximum power point (Pmax), and fill factor (FF) and other relevant parameters can be calculated. Connect to the photoelectrochemical module for measuring solar energy materials’ light – modulated AC impedance IMPS/IMVS.
Can connect to a WIFI module to achieve wireless connection between computers and instruments; can remotely control equipment in the laboratory through the network in the office.
Batch processing functions: Users can freely combine multiple measurement procedures and automatically perform them in sequence.
Open – source programs: Provide open – source programs, allowing users to write various programs through different programming languages, such as C++, Delphi, VB, etc., and more conveniently remotely control instruments through LabVIEW software.
Has 2 analog signal input channels and 1 analog signal output channel.
Technical Parameters
Power Supply | Single – phase AC 110 – 220V ± 10%, 50Hz, power consumption about 300W. |
Channels | 2, expandable to 16 channels. |
Maximum Current | ±1A standard, expandable to ±10A without a booster. |
Maximum Output Voltage | ±20V. |
Applied Potential Range | – 10V ~ +10V. |
Potentiostat Bandwidth | >500kHz. |
Applied Potential Precision | 0.2% or 2mV. |
Current Range | ±100pA ~ ±1A. |
Applied Current Resolution | 0.008% of the applied current range. |
Measured Current Resolution | 0.003% of the selected current range, minimum 3fA. |
Measured Potential Resolution | 0.0008% of the selected potential range, minimum 7nV. |
Cyclic Voltammetry and Linear Sweep Rate | 1μV/s ~ 10000V/s. |
Timing Method Sampling Frequency | 10μHz ~ 1MHz. |
Disturbance Signal Amplitude | 0.15mV ~ 2V, or 0.03% ~ 100% of the current range. |
Input Impedance | >1000Gohm//<10pF. |
Input Offset Current | <20pA. |
Dimensions | Approximately W470D360H140mm. |
Weight | Approximately 7.8kg. |