Functional Features
- Dual-center mixing and defoaming device.
- Centrifugal force from rotation and convection generated by revolution homogenize sample materials.
- Contact-free design without mixing blades to minimize contamination.
- Two operation modes: automatic and manual; 6 sets of operating parameters can be stored in automatic mode.
- External vacuum pump is required for this equipment.
Specifications
| Power Supply | Single-phase AC 220V, 50Hz, total power approx. 450 W |
| Mixing Method | Planetary centrifugal mixing |
| Standard Container | 300 ml ×1 (Φ77×H85 mm) |
| Max Processing Capacity | 150 mL×1; 300 g×1 (gross weight) |
| Working Principle | Planetary gravity mixing (Rotation / Revolution = fixed 0.3 ratio) for stirring and defoaming |
| Revolution | Speed-adjustable (200-2000 rpm); arbitrary input, accuracy: 1 rpm |
| Rotation | 0.3 times of revolution speed |
| Max Speed | 2000 rpm |
| Drive Mode | 400 W motor |
| Stored Parameter Sets | 6 groups |
| Balance Adjustment | Self-balancing (no counterweight adjustment required) |
| Setting Time | Three-segment programmable, each segment 200 s, total 600 s (accuracy 1 s). (Auto mode: to prevent jar over-heating & melting, max 150 s per segment for 1450-1800 rpm; max 50 s per segment above 1800 rpm.) (Manual mode: for speed>1800 rpm, single-run max 120 s; auto 120-second cooling countdown upon completion.) |
| Vacuum | External vacuum pump is required |
| Control System | Dual closed-loop PLC control, Chinese UI, touch-screen operation |
| Lubrication System | Cup-splash-proof structure, passive lubrication to avoid bearing wear |
| Main Alarm Functions | Unbalanced condition / Upper cover not closed / Overload |
| Main Safety Functions | Auto-stop upon fault; upper cover auto-lock during operation |
| Transmission Type | Gear drive |
| Equipment Dimensions | Approx. L300 mm × W440 mm × H360 mm (excluding protruding parts, subject to actual product) |
| Weight | Approx. 55 kg (subject to actual product) |